Instruction ManualHigh Temperature PyranometerCM 4

Seite 2 - IMPORTANT USER INFORMATION

9 1 GENERAL INFORMATION 1.1 INTRODUCTION The CM 4 High Temperature Pyranometer is an instrument for measuring solar or artificial light i

Seite 3

10 The CM 4 Pyranometer complies with specifications according to the ISO 9060 standard, as defined in the ‘Guide to meteorological Instrument

Seite 4

11 The thermopile surface is coated with black absorbent paint. Absorbed radiation is converted into heat which flows through the thermal res

Seite 5 - DECLARATION OF CONFORMITY

12 Temperature dependency of the sensitivity-3.000-2.000-1.0000.0001.0002.0003.000-40 -20 0 20 40 60 80 100 120 140 160Instrument temperature (

Seite 6

13 1.2.2 Spectral properties of the glass dome The spectral properties of a pyranometer are determined by the properties of the black absorb

Seite 7

14 1.2.3 Directional / Cosine response The measurement of the radiation falling on a plane surface (also called irradiance or radiative flux)

Seite 8

15 -6-4-202468100 1020304050607080min. cosine error %typical cosine error %max. cosine error %degrees[%] Figure 1.4: Th

Seite 9

16 0.960.970.980.991.001.010 500 1000 1500 2000 2500Irradiance [W/m²]Relative error [-] Figure 1.5: CM 4 non-linearity, sensitivity variat

Seite 10 - 1 GENERAL INFORMATION

17 2 LIST OF SPECIFICATIONS Spectral range: 300 to 2800 nm, 50% points Sensitivity: 7 µV/Wm-2 (nominal) Impedance: 500 to 2000 Ω Res

Seite 11

18 Temperature sensor: Pt-100 Construction: Receiver paint: Carbon Black Dome: Glass Desiccant: Silica gel Materials: Anodised alumin

Seite 12 - 1. GENERAL INFORMATION

1 IMPORTANT USER INFORMATION Reading this entire manual is recommended for full understanding of the use of this product. The exclamation m

Seite 13 - Temperature dependency

19 Figure 2.1: CM 4 Dimensions. SPECIFICATIONS

Seite 15

21 3 INSTALLATION Reading the installation instruction before installation is recommended for full understanding of the use of this product.

Seite 16

22 3.2.1 Outdoor installation When installed permanently, the pyranometer can be attached to its mounting platform by means of the holes that

Seite 17 - Relative error [-]

23 3.3 ELECTRICAL CONNECTION The CM 4 is provided with a special 10 m cable with six leads and a shield covered with a black sleeve. The co

Seite 18 - 2 LIST OF SPECIFICATIONS

24 Figure 3.1: Circuit diagram of the CM 4 Pyranometer and connection to readout equipment. It is evident that applicatio

Seite 19

25 For amplification of the pyranometer signal Kipp & Zonen recommends the 4-20 mA Signal Amplifier, available from Kipp & Zonen. Thi

Seite 21

27 4 OPERATION After completing the installation the pyranometer will be ready for operation. The irradiance value (E) can be simply compu

Seite 23

2 COPYRIGHT© 2010 KIPP & ZONEN All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or transm

Seite 24 - 3. INSTALLATION

29 5 MAINTENANCE Once installed the pyranometer needs little maintenance. The pyranometer dome must be kept clean and inspected regularly.

Seite 26

31 6 CALIBRATION 6.1 INITIAL CALIBRATION The ideal pyranometer should always have a constant ratio of voltage output to irradiance level

Seite 27

32 Accurate calibrations can also be done outdoors under clear conditions by comparison to a reference pyrheliometer. Many National Weather Se

Seite 28 - 4 OPERATION

33 6.3 CALIBRATION PROCEDURE AT KIPP & ZONEN 6.3.1 The facility The calibration facility at Kipp & Zonen consists of a good qualit

Seite 29

34 the lamp optics etc. Therefore the pyranometers are interchanged and the whole procedure is repeated. We get another pair of values: A'

Seite 30 - 5 MAINTENANCE

35 6.3.5 Traceability to World Radiometric Reference Working reference pyranometers are maintained at Kipp & Zonen. Each reference pyrano

Seite 32 - 6 CALIBRATION

37 7 FREQUENTLY ASKED QUESTIONS (FAQ’s) The most frequently asked questions are listed below. 1. Negative output during measurements?

Seite 34 - 6. CALIBRATION

3 CALIBRATION CERTIFICATE The calibration certificate supplied with the instrument is valid from the date of first use. Even though the cal

Seite 35 - (1/e) of several minutes

39 8 TROUBLE SHOOTING The following contains a procedure for checking the instrument in case it appears that it does not function as one c

Seite 37

41 9 PART NUMBERS / SPARE PARTS / OPTIONS Description Part no. Drying Cartridge kit 0356 111 (incl. Cartridge, Cover, Rubber Ring) S

Seite 39

43 APPENDIX I PYRANOMETER CLASSIFICATION ACCORDING TO WMO GUIDE 1996 Characteristics High quality Good quality Moderate quality ISO 9060 c

Seite 40 - 8 TROUBLE SHOOTING

44 Achievable uncertainty, 95 percent confidence level Hourly totals Daily totals 3% 2% 8% 5% 20% 10%

Seite 41

45 APPENDIX II PT-100 SPECIFICATIONS Temp.[°C] Resistance [Ω] Temp. [°C] Resistance[Ω] Temp. [°C] Resistance [Ω] -40 -39 -38 -37 -36 -35

Seite 42

46 Temp. [°C] Resistance [Ω] Temp. [°C] Resistance [Ω] Temp. [°C] Resistance [Ω] 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75

Seite 43

47 APPENDIX III LIST OF WORLD AND REGIONAL RADIATION CENTRES World Radiation Centres Davos (Switzerland) St. Petersburg (Russia) Region

Seite 45

4 DECLARATION OF CONFORMITY According to EC guideline 89/336/EEC We Kipp & Zonen B.V. Delftechpark 36 2628 XH Delft The N

Seite 46 - Resistance

49 APPENDIX IV RECALIBRATION SERVICE Pyranometers, UV-meters, Pyrgeometers & Sunshine duration sensors Kipp & Zonen solar radiat

Seite 48 - RADIATION CENTRES

51 Name : Company/Institute : Address : Postcode + City : Country : Phone : Fax : E-mail :  I would like to receive a price estimate

Seite 49

Our customer support remains at your disposal for any maintenance or repair, calibration, supplies and spares.Für Servicearbeiten und Kali

Seite 52 - Netherlands

7 TABLE OF CONTENTS IMPORTANT USER INFORMATION....1 CALIBRATION CERTIFICATE ...

Seite 53 - Passion for Precision

8 6.3.5 Traceability to World Radiometric Reference.....35 7 FREQUENTLY ASKED QUESTIONS (FAQ’S).......37 8 TROUBLE