11.3 Data Format
The data logger COMBILOG 1022 supports following data
formats:
For-
mat
Start-
bit
Data-
bit
Parity-
bit
Stop-
bit
Character
length
ASCII /
Modbus
Profibus
8N1 1 8 N 1 10 X
8E1 1 8 E 1 11 X X
8O1 1 8 O 1 11 X
8N2 1 8 N 2 11 X
8E2 1 8 E 2 12 X
8O2 1 8 O 2 12 X
Table 11.1 Supported data transfer formats
The data format 8E1 with even parity (E=even) corresponds to
the PROFIBUS-definitions according to DIN 19245, part 1, and
is supported by the data logger both in the PROFIBUS-protocol
and in the ASCII- and MODBUS-protocol. This data format
should therefore generally be selected for the transmission.
For modem connections, which in most cases can be carried
out without a parity-bit, the second data format is available.
This data format is only supported by the ASCII-protocol and
the MODBUS-protocol.
The data format is defined via configuration software
COMBILOG.EXE. If there are no other specifications, the de-
fault adjustment is ”no parity.