11.5 Logging options.
In the logging window the logging period can be selected in the range of 10 seconds up to
1hour.
Sensor start-up delay can be selected in the range from 0 seconds up to 9 seconds.
Memory management (for the internal memory of 128 kB) has the following options: circular
(after the memory is full, the oldest data will be over written) or on-stop (after all memory is
used, new data will no longer be stored in memory, even though the measurements will be
performed). Important: When using the SD card the logging to memory should be set to
Circular. Otherwise the logging stops after the (internal) memory is full.