39
This method works well if the instrument is on a vertical post attached to the boom extending
from the tower. The pyranometer is levelled while the post is vertical in an upright position.
The measurement of the angle of the post can be accomplished to within 0.1° using a high
quality carpenter’s level.
(2) The second procedure requires the construction of a levelling jig. This consists of a flat planed
parallel piece of metal attached to a circular ring whose diameter is such that it will sit around
the outside dome of the pyranometers to be inverted. The ring must have known parallel ends.
The metal flat (which can be reversed, side-to-side) is attached to one end of the ring, while
the other end of the ring sits on the ring surrounding the pyranometer outer dom e. To improve
the performance of this tool, three small ‘feet’ may extrude from the instrument end of the
ring for positive placement on the pyranometer. On the far end of the metal plate an adjustable
circular spirit level is attached for the ultimate levelling of the pyranometer (pyrgeometer)
to be used in the downfacing position.
The pyranometer is first levelled in its normal position following radiometric levelling of the
instrument. The levelling tool is placed on the pyranometer and the adjustable level on the
plate set to conform to the instrument bubble level. The plate is then turned over so that the
bubble level will be upright when the pyranometer is inverted.
When attaching the pyranometer to its inverted position, spring loaded retaining bolts are required
to maintain a constant pressure against which the levelling feet can be adjusted. The level can be set
by holding the levelling jig against the instrument and adjusting the levelling feet of the pyranometer
in the normal manner.
4.3 Installation of instruments for the measurement of direct beam radiation
4.3.1 General Considerations
The original goal of the BSRN was to use a cavity radiometer with an open entrance aperture for the
measurem ent of direct beam radiation. This was later amended to include the use of a norm al incidence
pyrheliometer (or more simply pyrheliometer) to fill gaps in the data stream during those time periods
when the cavity radiometer was in calibration mode. Further amendments were made to the original
concept when concerns about protecting the open cavity radiometer against the elements were brought
forward. Further investigations have found that calcium fluoride or sapphire optical flats provide good
protection from the elements while transmitting virtually the full solar spectrum. The ideal configuration
for the measurement of direct beam radiation remains the use of either an open all-weather cavity
radiometer or all-weather cavity radiometer with the appropriate optical flat used for protection along
with a pyrheliometer preferably capable of measuring the same spectral range, that can be used to
complete the data set during the periods the cavity radiometer is in calibration mode. Thus the cavity
radiometer is the primary instrument with the pyrheliometer being used to fill the ‘calibration gaps’ by
correlation with the observations obtained during the cavity measurem ents periods preceding and
following the calibration time period. The actual frequency and length of time required for the self-calibration
period depends upon the type of the cavity radiometer. Lesser alternatives, however, are acceptable.
In rank order of preference these are: (1) The use of a pyrheliometer as the primary instrument while
an open cavity radiometer is used in tandem at all times weather conditions permit. In this manner,
the pyrheliometer is calibrated against the cavity radiometer nearly continuously. (2) same as (1), but
with a cavity radiometer with a quartz flat covering the entrance aperture. This cavity in turn is to be
calibrated against an open aperture cavity radiometer to account for the effect of the flat. (3) The use
of two pyrheliometers measuring on a routine base, with an open aperture cavity radiometer checking
the calibration on a periodic basis during high solar radiation conditions.
4.3.2 Pre-installation checks and service
(1) If not provided by the manufacturer, the instrument should be calibrated so that the following
information is available:
(i) the responsivity of the instrument to radiation
(ii) the linearity of the instrument between 0 and 1500 Wm
-2
Kommentare zu diesen Handbüchern